HFIL?
High Level Setpoint Query
Input: HFIL?
Returned:
XXX.X
Comments:
Used to query the High Level Setpoint. Value returned in either
inches or millimeters depending on the current level units. There must be at least
a 0.5 inch interval between each level setting and alarm.
HILA
Set High Level Alarm
Input: HILA XXX.X
Returned:
No response
Comments:
Used to set the High Level Alarm value. Input value uses the
current level units, either inches or millimeters. There must be at least a 0.5 inch
interval between each level setting and alarm.
HILA?
High Level Alarm Query
Input: HILA?
Returned:
XXX.X
Comments:
Used to query the High Level Alarm value. Value returned in either
inches or millimeters depending on the current level units. There must be at least
a 0.5 inch interval between each level setting and alarm.
HILS?
High Level Alarm Status Query
Input: HILS?
Returned:
0 or 1
Comments:
Used to query the High Level Alarm Status where 0 = off and 1 =
on.
HILM
Set High Level Alarm Mask
Input: HILM 0 or HILM 1
Returned:
No response
Comments:
Used to enable or disable the High Level Alarm where 0 = disabled
and 1 = enabled. Default setting is 1 = enabled.
HILM?
High Level Alarm Mask Query
Input: HILM?
Returned:
0 or 1
Comments:
Used to query the High Level Alarm Mask status where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
LEVL?
Current LN2 Level Query
Input: LEVEL?
Returned:
XXX.X
Comments:
Used to query the current LN2 Level reading.
LFIL
Set Low Level Setpoint
Input: LFIL XXX.X
Returned:
No response
Comments:
Used to set the Low Level Setpoint. Input value uses the current
level units, either inches or millimeters. There must be at least a 0.5 inch interval
between each level setting and alarm.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...