11. Wait 180 seconds
After the controller counts down, the display will read
“Place Probe A in ice water”. Completely submerge the
sensing end of the probe in the ice water bath.
12. Press “ENTER”
The display will read “Wait for Temp A to stabilize”. Wait
for the displayed Temp A reading to stabilize while the
probe is submerged in the ice bath.
13. Press “ENTER”
The display will read “Probe A two point calibration
complete”.
Place Probe A in
ice water and press
ENTER
Wait for Temp A to
stabilize then press
ENTER
Temp A 0 °C
Probe A two point
calibration complete