MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
158
EVNLOG?
Event Log Record Query
Input: EVNLOG? N
Returned:
XXXXX, MM/DD/YY, HH:MM, XXX.X, XXX.X, XXX.X, XXX.X
Comments:
Used to query Event Record [n] where n = record number. String
returned: Date, Time, TempA, Temp B, LN2 Level, Liquid Usage, Event Codes.
LOGPER
Set Event Log Interval
Input: LOGPER XXX
Returned:
No response
Comments:
Used to set the event log interval between 1 and 240 minutes.
Default setting is 240 minutes.
LOGPER?
Event Log Interval Query
Input: LOGPER?
Returned:
XXX
Comments:
Used to query the current event log interval.
TIME
Set the Time
Input: hh:mm:ss
Returned:
No response
Comments:
Used to set the controller Time in the format hh:mm:ss
(hour:minute:second).
TIME?
Time Query
Input: TIME?
Returned:
hh:mm:ss
Comments:
Used to query the current time.
UNID
Set Unit ID
Input: UNID XXXXX
Returned:
No response
Comments:
Used to set the controller 5-digit Unit ID. TEC 3000 Unit ID should
be between 1 and 200.
UNID?
Unit ID Query
Input: UNID?
Returned:
XXXXX
Comments:
Used to query the controller Unit ID.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...