MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
66
7.2.5.2. Maximum Fill Time
The maximum fill time can be adjusted in the Advanced Fill Menus. If a freezer begins
filling but does not reach its High Level Setpoint within the maximum fill time, then the
TEC 3000 will terminate the fill and trigger a Fill Time Alarm. Auto Fill Control will then be
disabled until the Fill Time Alarm is cleared. The maximum fill time has a range of 30 to
240 minutes.
NOTE: Security Level 3 or higher is required to adjust the maximum fill time.
1. Press “SETUP”
Controller will prompt for a password. A flashing cursor
on the entry screen will make it clear which digit is being
changed. Use the “▲/▼” keys to scroll to the
appr
opriate number. Press “ENTER” to advance the
cursor to the next position.
2. Press “ENTER”
The display will read “Temperature Menus”.
3. Press “SETUP”
Until the display reads “Advanced Settings”.
4. Press “ENTER”
The display will read “Advanced Fill Menus”.
5. Press “ENTER”
The display will read “Timed Fill”.
6. Press “SETUP”
Until the display reads “Maximum Fill Time”. Use the
“▲/▼” keys to adjust and press “ENTER” to save.
Higher User Level
Required use ▲▼ to
ENTER Password
0000
Press ENTER for
Temperature Menus
or press SETUP for
next menu
Press ENTER for
Advanced Settings
or press SETUP for
next menu
Press ENTER for
Advanced Fill menus
or press SETUP for
next menu
Timed Fill
DISABLED
Use ▲▼ to adjust
Press ENTER to save
Maximum Fill Time
60 min
Use ▲▼ to adjust
Press ENTER to save
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...