HITSTB
Temp B High Alarm Test
Input: HITSTB
Returned:
No response
Comments:
Used to initiate the High Temp B Alarm Test. See the High
Temperature Alarm Test section for more information.
INITEE
Restore All Defaults
Input: INITEE
Returned:
No response
Comments:
Used to restore all factory default settings.
LNSATP
Set LN2 Saturation Temperature
Input: LNSATP ±XXX.X
Returned:
No response
Comments:
Used to set the LN2 saturation temperature for single point
calibration reference point.
Temperature Commands
BPTMP?
Bypass Temp Query
Input: BPTMP?
Returned:
±XXX.X
Comments:
Used to query the current bypass temp sensor reading.
HITA
Set Temp A High Alarm
Input: HITA ±XXX.X
Returned:
No response
Comments:
Used to set the Temp A High Alarm value.
HITA?
Temp A High Alarm Query
Input: HITA?
Returned:
±XXX.X
Comments:
Used to query the current Temp A High Alarm value
HITAS?
Temp A High Alarm Status Query
Input: HITAS?
Returned:
0 or 1
Comments:
Used to query the Temp A High Alarm status where 0 = Off
and 1 = On.
HITAM
Set Temp A High Alarm Mask
Input: HITAM 0 or HITAM 1
Returned:
No response
Comments:
Used to enable or disable the Temp A High Alarm where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
HITAM?
Temp A High Alarm Mask Query
Input: HITAM?
Returned:
0 or 1
Comments:
Used to query the Temp A High Alarm Mask status where 0 =
disabled and 1 = enabled. Default setting is 1 = enabled.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...