MVE Biological Systems
TEC 3000 Operating and Maintenance Freezer Manual
13289499 Rev H •
94
Installing MVE TEC COM USB Kit
1. Insert the included USB to Serial Driver Disc into the PC CD-ROM
2. Plug the MVE TEC COM USB Kit into an open USB port
3. Follow the automatic installation prompts
Downloading TEC Connect
1. Uninstall in any previous TEC Connect versions
2. The latest version is available online at the Chart Industries website
3. Click the link and open the compressed folder
4. Extract all files from the compressed folder to a location on your hard drive
5. Open the extracted SETUP file
6. Follow the prompts to complete installation
7. A TEC Connect icon should appear on your programs list when complete
Connecting to Chart TS
Chart TS Settings
1.
COM Setup: “9600 N81”
2. COM Type:
“ASCII”
3. Select a unique MODBUS ID (Unit ID) for each controller involved
TEC Connect Settings
1. Connect Chart TS serial port to PC with MVE TEC COM USB Kit.
2. Open TEC Connect
3. The TEC Connect program Determine the Windows COM Port. This can be
accomplished by navigating to the device manager and locating the COM port
labeled “RS-485 Isolated Port”
4. Note the COM number listed
beside “RS-485 Isolated Port” in the Device Manager
window.
5. Close Device Manager and click
“OK”
6. Select the appropriate Windows COM Port from the drop down menu
7. Connect to Chart TS
•
For a single controller:
Check the “Auto Find” box or enter the
MODBUS ID of the controller and click
“Connect”.
•
For networked controllers:
Uncheck the “Auto Find” box, enter the
Unit ID of the desired controller, and click
“Connect”. In order to connect to another
controller on the network, click “Disconnect”,
change the Unit ID,
and then click “Connect”.
8. If the controller is successfully connected, the current temperature and level
information from the controller will be displayed and “COMX connected” will be
displayed under “Status”, where X is the COM port the TEC COM kit is connected
to on the PC.
NOTE: Contact your MVE Distributor or Technical Service if you experience
problems connecting to a controller.
Downloading the Event Log
1. Connect to controller using TEC Connect
2. Select Event Download Range with Event 1 being the most recent logged event.
The default download range includes all events.
3.
Select the “Download” button. Once the event log has finished downloading, select
“Open Download” button.
Содержание MVE 1842P-150
Страница 1: ...Chart Cryogenic Freezers with MVE TEC3000 Controllers Operating and Maintenance Manual ...
Страница 24: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 22 11 10 ...
Страница 106: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 105 8 4 Add On Menus ...
Страница 109: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 108 8 7 Liquid Level Menus ...
Страница 111: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 110 8 9 Advanced Settings Menus ...
Страница 112: ...MVE Biological Systems TEC 3000 Operating and Maintenance Freezer Manual 13289499 Rev H 111 8 10 Password Menus ...