2. Installation procedures
-
37
-
●
For normal imaging
X-ray exposure
signal (RX_REQ)
X-ray release signal
(RX_COM)
Actual X-ray exposure time
<
300ms
Edge detection and status detection are used together
to detect the X-ray exposure signal (RX_REQ).
Imaging time:
Normally up to
1100 ms
Long periods of time:
Up to 3100ms
< 100ms
< 100ms
T
PRE_DELAY
T
RX_REQ_NEG
T
PREP
●
When RX_REQ is not negated, and a timeout occurs
X-ray exposure
signal (RX_REQ)
X-ray release signal
(RX_COM)
Actual X-ray exposure time
<
300ms
Timeout: Normally 1100 ms
Long periods of time: Up to 3100ms
< 100ms
< 100ms
T
PRE_DELAY
T
POST_DELAY
T
PREP