2. Installation procedures
- 19 -
(2) Wired connection (option)
No.
Item
Conditions and checkpoints
Reference items
1
Installing the driver of
infrared data
communication unit,
• The driver installation must be completed
successfully.
Refer to “7.1.2
Connecting the
infrared data
communication unit”.
2
Connecting the infrared
data communication unit
with the image capture
computer
• Connect the unit to the USB port on the image capture
computer.
• Turn on the power of the image capture computer and
start the link controller.
Refer to “7.1.2
Connecting the
infrared data
communication unit”.
3
Turning on the power of
the detector
• A fully charged battery must be installed properly in
the detector.
• The LED on the detector must light.
Refer to “7.1.6
Registration of
detector”.
4
Registration of detector
• The distance between the detector and infrared data
communication unit must be less than 30 cm.
• Care must be taken not to drop the detector.
• The detector must be registered in a location outside
the patient environment.
Refer to “7.1.6
Registration of
detector”.
5
Connecting the detector
and AC/DC box using the
sensor cable
• Check the directions in which the cable connectors
are inserted, and plug them in.
Refer to “7.1.8 When
connecting the wiring
unit”.
6
Connecting the status
indicator to the AC/DC
box
• Check the directions in which the cable connectors
are inserted, and plug them in.
Refer to “7.1.8 When
connecting the wiring
unit”.
7
Connecting the image
capture computer with
the AC/DC box
• This connection must be made through a hub.
• Route the LAN cable in such a way that no excessive
weight will be applied to it.
Refer to “7.1.8 When
connecting the wiring
unit”.
8
Turning on the power of
the AC/DC box
• The battery must be installed properly in the detector.
• The LED on the detector must light.
9
Connecting the image
capture computer with
the X-ray interface box
• This connection must be made through a hub.
• Route the LAN cable in such a way that no excessive
weight will be applied to it.
Refer to “7.1.7
Connecting to the
X-ray interface box”.
10
Connecting the X-ray
generator to the X-ray
interface box
• Route the X-ray interface cable in such a way that no
excessive weight will be applied to it.
Refer to “7.1.7
Connecting to the
X-ray interface box”.
11
Establishing the CXDI
Control Software NE
settings
• Set the items which are required.
Refer to “7.4 CXDI
Control Software NE
settings”.
12
Connecting to the
network and setting the
output destination
Refer to “7.4 CXDI
Control Software NE
settings”.
13
Aligning the detector and
X-ray tube exposure
fields
• The center positions of the detector and X-ray tube
must be aligned.
Refer to “7.5 Image
Quality Checks”.
14
Calibration
• No errors must be displayed.
Refer to “3.4 QC Tool
Button” in CXDI
Control Software NE
Setup guide.
15
Self-diagnosis test
• No errors must be displayed.
Refer to “3.4 QC Tool
Button” in CXDI
Control Software NE
Setup guide.
16 Imaging
test
• Transfer images to the printer and storage unit, and
check their image quality.
Refer to “7.5 Image
Quality Checks”.
17
Final check
• Carry out the final checks as per the check sheet.
Refer to “7.6
Post-installation
checks”.
18 Cleaning
* When performing both wireless connections and wired connections for one detector, first proceed
with installation items No.1 to No.5 for the wireless connections.
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...