2. Installation procedures
- 24 -
3) Perform the detector registration by tapping
the power switch on the side of the detector.
• At this point in time, the LED (blue-green) of the
detector blinks for an instant.
• The infrared data communication unit communicates
with the infrared communication area of the detector.
[When using a wired connection]
When the detector is registered properly, the
three LEDs (blue, green and blue-green) light up
simultaneously for about 1 second. After this, the
communication circuit is established, and only the
green LED goes off.
[When using a wireless connection]
When the detector is registered properly, the three LEDs (blue, green and blue-green) light up
simultaneously for about 1 second, and then operation moves to step 4).
Note:
* The distance between the signal sensor of the infrared data communication unit and the infrared
communication area on the detector must be less than 30 cm and the incident angle of the signals must
be less than 10 degrees. The work environment must not be close to the monitor or an environment
where signals are reflected.
* The LED (blue-green) on the detector does not light when the detector is not registered or when the
communication circuit is not established.
4) Connection to the access point is made automatically, and the communication circuit is established.
• Two LEDs (blue and blue-green) on the detector light up. (The green LED goes off.)
Note:
* If the communication circuit is not established, the blue-green LED on the detector will not light.
* Do not place any structures that will block communication between the detector and access point.
5) Start the service tool, and check that the detector which has registered the detector is displayed.
Select [Service Tool]
→
[Connection]
→
[Detector], and then open the [Ferry Static Detector] screen.
• If, after the detector has been registered, the detector is not reflected on the Connection tab of the
service tool, start the service tool again.
• For details, refer to “4. Essential Settings” in Chapter 2. Installation Manual of the CXDI Control
Software NE V.1.40 Service Manual.
(3) When using wireless and wired connection with one detector
Infrared communication area
Infrared data communication unit
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...