2. Installation procedures
- 22 -
7.1.3 Connecting the wireless LAN access point
1) Connect the image capture computer and access point through the hub.
• The hub/LAN cable to be procured from the distributor must support Gigabit Ethernet.
2) Turn on the power of access point.
7.1.4 Establishing the wireless LAN access point settings
1) Turn on the power of the image capture computer, start the access point setting tool, and establish the
access point settings (wireless communication settings). The access point setting tool is started from
the service tools.
• How to start the access point setting tool
Start ServiceTool.exe, and select [Tool]
→
[AP/LNK configuration tool] to start the tool.
• For details, refer to “4.2.2. For a Ferry protocol- and wireless-connected detector” in “4. Essential
Settings” in Chapter 2. Installation Manual of the CXDI Control Software NE V.1.40 Service
Manual.
7.1.5 Installing the wireless LAN access point
1) Install the wireless LAN access point in such as way that, using the access point as the starting point,
the distance between the access point and detector is less than 8 meters.
• Do not place any structures that will block communication between the detector and access point.
Detector
Access point
Hub
Operating range: Max. 8 meters
(starting from the access point)
Access point
LAN cable
Image
Capture
Computer
Hub
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...