2. Installation procedures
- 1 -
1. Work-related precautions
Bear the following points in mind when installing the system.
(1) Carry out the work while ensuring that the none of the portable units will be dropped or subjected to
impact.
(2) Ensure that the installation location satisfies the following criteria:
1) A location where the equipment will not be splashed with water
2) A location where the air does not contain moisture, dust, salt, sulfur or other such substances and
where the equipment will not be adversely affected by poor ventilation, the air pressure level, the
temperature and other such factors
3) A location which is not tilted or subject to vibration or impact (this includes the locations when the
equipment is conveyed)
4) A location where chemical products are not stored and where no gases are given off
5) A location where the power source providing the correct power line frequency, voltage and power
used by the equipment can be supplied
6) A location where a grounding wire and the 3-pin grounding plug of the power cable can be
connected
(3) After installing the equipment, ensure that the product packaging materials and other waste materials
generated at the time of installation are disposed of in an environmentally sound manner.
2. Installation restrictions
(1) Take steps to ensure that the system is not installed near electronic devices or equipment since noise
artifacts occur more readily on the images in environments where strong electromagnetic fields are
present.
Examples of electronic devices and equipment concerned: CRT monitors, X-ray generators, medical
electronic equipment
(2) The system must not be installed near devices or equipment which may interfere with wireless
communication.
Examples of electronic devices and equipment concerned: Thermal therapy devices, laser knives,
cordless telephones
(3) When using a wired connection for the product, install the detector and AC/DC box with a distance of
at least 150 mm between them.
(4) For use, the cables must be routed without bending, pinching or stretching them or subjecting them to
unreasonable loads. Remodeling of the cables is prohibited.
(5) In the case of a network connection, the appropriate network address must be acquired from the
network administrator in the installation location, and it must be set correctly.
(6) The wireless LAN access point, switching hub, battery charger, image capture computer, infrared
data communication unit and X-ray interface box must be installed outside the patient environment.
For further details, refer to section “2. System diagram” in “Chapter 1 Overview.”
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...