1
1
.
.
O
O
v
v
e
e
r
r
v
v
i
i
e
e
w
w
- 10 -
5.5 Required PC specifications
Use of the CXDI-80C Wireless requires a PC (hereinafter image capture computer) running
Microsoft Windows 7 and on which CXDI Control Software NE (Ver. 1.40 or later) or CXDI
Controller RF (Ver. 1.40 or later) made by Canon has been installed, and hardware that satisfies
the required specifications must be procured from a Canon sales company or agent. For details
of the required specifications of the image capture computer, refer to the CXDI Control
Software NE (Ver. 1.40 or later) or CXDI Controller RF (Ver. 1.40 or later) service manual.
5.6 Wireless communication specifications
The relay unit used for wireless communication between the CXDI-80C Wireless and Image
Capture Computer is called a wireless LAN access point. The table below lists the wireless
communication specifications.
Item
Details
Remarks
Communication
system
IEEE standard
IEEE 802.11n
Frequency band
5 GHz/2.4 GHz
Security
(encryption
system)
WPA2-PSK
(AES)
AES is an encryption
system which has been
certified by the
National Institute of
Standards and
Technology (NIST) of
the United States.
Channel
bonding
ON (40 MHz)
(Recommended)
Channel settings are
used to avoid
interference in
adjacent imaging
rooms.
Communication
environment
Communication
range
Within 8 meters
Recommended in
order to meet detector
specifications.
Ambient
conditions
•
Do not place any structures that
will block the wireless signals
between components of the system.
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...