2. Installation procedures
-
42
-
7.4 CXDI Control Software NE settings
The various settings are established using CXDI Control Software NE. For details on these
settings, refer to the “CXDI Control Software NE Service Manual.”
[Reference information]
(1) IP addresses of image capture computer (System IP addresses)
Item
Setting value
Local IP address
192.168.100.10
Subnet mask
255.255.255.0
(2) IP addresses allocated to 80C detector (PCA-FE board) (Target IP addresses)
Item
Setting value
80C detector IP address
192.168.100.11
80C detector IP address*
192.168.100.12–
(digits in fourth column
changed)
Subnet mask
255.255.255.0
* When using wireless and wired connections with one detector, allocate different IP
addresses.
* When using two detectors in the same room, allocate different IP addresses.
* When bringing in a detector from one room to be used in another room, allocate different
IP addresses.
(3) IP addresses registered in X-ray interface box (PCA-XIF board) (factory defaults)
Item
Default
X-ray interface box IP address
192.168.100.96
Subnet mask
255.255.255.0
(4) IP addresses registered in existing detector (PWB-Di board) (factory defaults)
Item
Default
Existing detector IP address*
192.168.100.11
Subnet mask
255.255.255.0
* Existing detectors: CXDI-40EG/EC, 40G compact, 50G/C, 60G/C, 55G/C
* For details on changing the IP addresses of existing detectors, refer to the “Tool Software
Operation Manual.”
(5) Settings which must be established
Upon completion of the system connections, the following settings must be established. For
details, see the CXDI Control Software NE Service Manual 2. Installation Manual.
• Create a predetermined Workspace and associate it with the detector.
• Perform calibration, performance tests and self-diagnosis.
• Adjust the photo timer
Note: The EI value used in the photo timer adjustment is 280.
(6) Settings
A number of other settings are required depending on the configuration of the equipment at
the location where the system is installed and on the requirements of the user.
• User settings
• Protocol creation
• RIS, PACS, printer connection settings
• Other necessary items
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...