2. Installation procedures
- 27 -
(2) Cable connections
1) Remove the skin-top spiral piece and lock nut of the X-ray interface cable.
2) As shown in the figure below, pass the X-ray interface cable through the X-ray interface box cabinet,
and loosely tighten the lock nut.
3) Pass the X-ray interface cable through the cable clamp, and connect the cable connector to the X-ray
interface PCB. Do not subject the cable to any stress and also, in order to prevent the excess length
from touching the devices on the circuit board, adjust the cable length and tighten the skin-top spiral
piece in such a way that no gap will be left as shown in the figure below.
4) Using a wrench, fasten the skin-top spiral piece and lock nut which were loosely tightened.
Skin-top spiral piece
Lock nut
Cable clamp
Connectors
No gap to be
left here
Skin-top spiral piece
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...