4. Repair Guide
- 16 -
After updating has been completed in the backup mode boot status, follow the steps below for
booting in the normal mode.
1) Hold down the power switch of the detector for at least five seconds to turn the unit’s power
OFF.
2) With a wired connection using the wiring unit, turn off the power of the AC/DC box.
3) Remove the battery.
4) Return the boot switch to the normal mode setting.
5) Connect a battery with a full charge again.
6) With a wired connection using the wiring unit, turn on the power of the AC/DC box.
7) Press the power switch of the detector for at least a second to turn on the unit’s power.
When the power is turned on with the boot switch at the normal mode setting, the normal code
firmware is loaded into the SDRAM, and operations are performed.
* When booting the normal mode after the backup mode booting, check the versions of the
firmware and FPGA from the Ferry type FPD maintenance tool and perform update to
have the right combination of them.
Boost switch
Battery holder
Normal mode
Backup mode
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...