4. Repair Guide
- 23 -
5.3 Other reference information
5.3.1 Changing the IP address of the PCA-XIF board
The IP address of the PCA-XIF board can be changed or checked using the “ipadr” command in the
telnet connection status.
Example 1: When changing the IP address from “192.168.100.96” to “192.168.100.97”
XIF_GUEST>ipadr
∆
-ip
∆
192.168.100.97 (where
∆
is a half-size space)
* The power must be turned off and turned back on for the new setting to be reflected.
* To set the IP address of the PCA-XIF board into the detector, start the AP/LNK
configuration tool of Service Tool, and take the following steps.
1)
Input the IP address (192.168.100.97 in this case) of the PCA-XIF board which has
been changed to the X-ray interface address on the Configuration tab screen, and
press the “Save” button.
2)
Bring the infrared communication area of the detector close to the infrared
communication unit, and proceed with the LINK operation.
(Commentary)
After “ipadr-ip 192.168.100.97” has been input, the “New setting takes effect after power cycling”
message is displayed. Before turning the power back on, input “ipadr,” and when the current status is
checked, “192.168.100.96” will be displayed for Current and “192.168.100.97” for the flash memory.
At this point in time, it is clear that only “192.168.100.97” has been written in the flash memory and that
the new setting has not been reflected.
Screen on which to change and check the IP addresses
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...