2. Installation procedures
- 30 -
7.1.8 When connecting the wiring unit (option)
The detector has an external connector which enables the wireless connections when wiring unit is
connected to it.
The wiring unit is used to establish wired connections, and it supports detection while it remains on an
upright stand or recumbent table Bucky unit.
By establishing the wired connection, communication is performed and power is supplied along the
detector cable so this can be used as a backup connection when the communication circuit with a
wireless connection is unstable or when the battery charge level is insufficient. When establishing both
wireless connections and wired connections for one wireless detector, first carry out the steps in sections
7.1.2 to 7.1.7, and then follow the instructions given in this section.
When building a system solely with wired connections using the wiring unit, the steps in sections 7.1.3
to 7.1.5 need not be taken.
[Checkpoints]
• The battery must be installed in the detector even when wired connections are established.
• The battery will not be recharged through the wiring unit. The battery is only be charged
through the battery charger.
[Switching between wireless and wired connections]
This switching is done automatically by detecting whether the detector cable is connected or not.
However, since the IP addresses allocated to the detector differ for wireless connection and for wired
connection, the detector must be selected again in the CXDI control software.
(1) Cable connections
1) Connect the AC power cable, status indicator cable, LAN cable and sensor cable to the connector
connection area on the AC/DC box.
* Be absolutely sure to use the AC power cable which comes supplied as an accessory with this
product.
* Check the directions in which the sensor cable connectors are to be connected before plugging
in the connectors.
* Check the directions in which the status indicator cable connectors are to be connected before
plugging in the connectors.
* Route the cable without forcibly bending, pinching or stretching them.
2) Connect the LAN cable, which has been connected to the AC/DC box, to the hub.
* The hub must be connected to the Image capture computer.
LAN cable
Detector cable
AC power cable
Status indicator cable
AC/DC box
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...