4. Repair Guide
- 6 -
3.2.2.2 Mounting the PCA-XIF board
1) Insert the LAN cable connector, AC adapter connector (DC inlet) and LED of the PCA-XIF
board through the holes in the base, and lay the PCA-XIF board flat.
* The shielded plate that covers the LAN cable connector is made electrically conductive
with the edge by pushing it into the hole in the base.
2) Using the DC inlet position as reference, adjust the position of the PCA-XIF board.
* The end of the DC inlet must protrude by about 0.1 mm from the base wall.
3) Secure the PCA-XIF board using the eight screws (XB1-2300-606).
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
PCA-XIF board
X-ray interface box base
DC inlet
Wall of X-ray interface box base
Protrusion amount: Approx. 0.1 mm
XB1-2300-606 screws (x 8)
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...