3
3
.
.
Description of Functions
- 8 -
2.2.1 PCA-XIF board
The PCA-XIF board is located inside the X-ray interface box, and has the main functions which
are described below.
(1) It has a function for interfacing with X-ray generator. In addition, the board employs a relay
circuit for the output signals and a photo-coupler for the input signals to protect the internal
circuitry.
(2) It has an exposure control function which monitors the statuses of the X-ray generator and
detector to determine whether exposure is possible and, if it is possible, to enable it.
(3) It has a function for interfacing with the detector.
(4) It uses LEDs to display the power supply status.
* The dedicated AC adapter is used to supply power to the PCA-XIF board. It does not have a
power switch because the X-ray interface box and the AC adapter are supported to be used by
continuous energization. The PCA-XIF board must be turned on and off by connecting or
disconnecting the plug on the wall.
* For safety concerns, do not disconnect the cable from the X-ray interface box or disconnect
the AC adaptor. It would expose the electrically-activated part. To avoid this, please
disconnect the plug on the wall or turn off the commercial power supply.
* The use of table tap, which causes a change in electrical safety, is not allowed according to the
electrical safety standards.
* The X-ray interface box must be placed outside the patient environment.
SW1 (RESET) switch
SW2 (BOOTSEL) switch
X-ray interface cable
LAN cable connector
AC adapter connector
LED
Grounding terminal
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...