4. Repair Guide
- 7 -
3.3 Wiring unit
Turn the power of the AC/DC box off, and disconnect the AC power cable, detector cable,
status indicator unit and LAN cable. Do not pull the cables free in order to disconnect the
connectors. When re-assembling the parts, follow the procedure for their disassembly in reverse
unless otherwise specified.
3.3.1 Removing the top cover
Loosen the two screws (XB2-7300-606) on the rear panel of the AC/DC box, remove the four
M3L6 brazier head screws (XA9-2033) on the side panels, and then remove the top cover of the
AC/DC box.
* Take care not to deform the top cover which has been removed.
* When re-attaching the top cover, tighten up the screws in the following sequence:
connector panel (screws with external teeth)
→
both side panels (brazier head screws).
Reference:
Tightening torque: 78.4 N-cm (8 kgf-cm) for side panel M3L6 brazier head screws
Tightening torque: 117.6 N-cm (12 kgf-cm) for bottom panel screws
These screws need only be loosened.
Top cover of AC/DC box
M3L6 brazier head screws
(XA9-2033) (x 4)
XB2-7300-606 screws (x 2)
AC/DC box base
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...