3.3.4 Replacing the PCA-PB board
3.3.4.1 Removing the PCA-PB board
1) Remove the four screws (XB1-2300-606) securing the PCA-PB board.
2) Slide the PCA-PB board in the direction of the red arrow shown in the figure, draw the
detector cable connector part out through the hole in the base, and remove the board.
3.3.4.2 Mounting the PCA-PB board
As a basic rule, mount the PCA-PB board by following the procedure for its removal in reverse.
Pass the detector connector part of the PCA-PB board through the hole in the base, and secure
the PCA-PB board using the four screws (XB1-2300-606).
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
XB1-2300-606 screws (x 4)
PCA-PB board
AC/DC box base
Draw the detector cable connector part out
from the hole in the base.
When the board is assembled, ensure that the detector cable
connector part protrudes from the base by about 4.5 mm.