2. Installation procedures
- 2 -
(7) There are no usage restrictions on how the detector is to be positioned.
(8) Avoid using the detector where it will be subjected to extreme bending loads or where it will easily be
dropped.
(9) Only one X-ray interface box can be connected to one CXDI-80C system. In the case of double X-ray
tube system, the connection with the X-ray interface box must be switched at the X-ray generator
side.
(10) Note that the radio frequency channel configured for indoor use may not be usable in outdoor areas
in the 5 GHz band. For further details, refer to section “5.6 Wireless communication specifications” in
“Chapter 1 Overview.”
3. Installation-related precautions
(1) Bear in mind that there will be a risk of malfunctioning and other trouble if this product is used
together with any equipment which may generates power supply noise, supply voltage fluctuations,
electromagnetic waves and other disturbances.
(2) Before proceeding with the installation work, ensure that the static charged in the bodies of the
installation personnel has been released.
(3) Before connecting or disconnecting the cables, be absolutely sure to turn off the power of the units in
order to ensure safety. In addition, do not touch the power outlets or connectors with wet hands.
* Checking out the wireless environment
1)
Make sure to perform the wireless LAN channel setting according to the channel setting guidelines.
2)
If the problem occurs on wireless communication, refer the Chapter 6. Troubleshooting [When
wireless communication trouble has occurred] to identify the cause and take necessary
countermeasures to solve the problem.
3)
If the signal strength level is still 0 or if it takes so much time to transfer the images even after
taking the countermeasures in step 2) above, explain the situation to the customer and ask them to
consider the use of wiring unit.
Содержание CXDI-80C
Страница 6: ...Note These measurements are only guidelines ...
Страница 10: ...1 1 O Ov ve er rv vi ie ew w 2 2 System diagram ...
Страница 48: ...2 Installation procedures 15 4 Handle unit option Corner pads Outer box Inner box ...
Страница 53: ...2 Installation procedures 20 7 Installation 7 1 Connections to the units 7 1 1 System connection diagram ...
Страница 82: ...2 Installation procedures 49 4 Battery Weight 0 25 kg Dimensional unit mm Dimensional tolerance 1 Scale 1 N ...
Страница 133: ...Rev 01 CONTENTS Detector unit 1 X ray interface box unit 2 Wiring unit 4 Handle unit 6 ...
Страница 134: ...Rev 01 Detector unit KEY NO PARTS NO Q TY DESCRIPTION MEMO 1 BA5 1230 000 2 Battery Connector Pad 1 1 ...