6. Troubleshooting
- 8 -
[No X-ray exposure possible]
Suspected causes:
•
Wireless communication trouble
*6
•
Incorrect X-ray interface IP address setting established by AP/LNK configuration tool
•
Improper connection of X-ray interface cable
•
Defective PCA-XIF board (inside the X-ray interface box)
Conduct checks based on the suspected causes listed above or replace the unit.
*6
If there is interference with signals of another wireless LAN device, it may be a case where the
RX_REQ signal from the X-ray generator cannot be recognized and X-ray exposure is disabled.
In such cases, use a wired connection with the wiring unit, and if X-ray exposure is enabled, it
means that trouble in the wireless communication is responsible for the problem.