3. CXDI-31 Imaging Unit
- 17 -
3.2.2.2.
Replacing the former sensor cable with the new sensor cable (BG7-3055)
Follow the steps below to replace the old type sensor cable (BG7-2356 or Y67-2676) with the
new type sensor cable (BG7-3055). The cable replacement requires the following parts in set,
which are listed in the Key no. 20, 32, 33, 34 in the [Table.1] below.
[Fig.1]
Parts list
Key No.
Description
Former Part No.
New Part No. Q’ty
Remark
BG7-2356-000
Without CLAMP
20
Cable Unit, Imaging
(Sensor)
Y67-2676-000
With CLAMP
BG7-3055-000
1 With
CLAMP
32 Clamp,
Cable
BA4-0567-020 BA4-2364-000
1
33 Screw
XA1-1260-406 XA1-1260-606
2 M2.6x6mm
34 Washer
-
XD1-1102-625
2 Newly
Added
[Table.1]
[Fig.2]
Old type sensor cable
(BG7-2356 or Y67-2676)
20-3
20-2
20-1
20
32
34
33
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......