2 Installation Manual
- 53 -
(5) -I
-I
A-RELEASE-RES is ignored.
Meaning
* This parameter is used simply as “-I”
Description * This is used when the error message “30012 Peer aborted Association (or never
connected)” occurseven though the DICOM data transfer was successful.
Æ
This is used based on the connected storage devices.
(6) -d FAC
-d FAC
This parameter dumps a specific facility log.
(DCM/DUL/SRV)
Meaning
* This parameter is used simply as “-d”
* This parameter is used to make the transfer software display the debugging
character string on the console.
Description * This parameter does not affect DICOM data transfer.
* This parameter outputs the CXDI log based on Windows NT.
(7) -v
-v
This parameter dumps the transfer log.
Meaning
* This parameter is used simply as “-v”.
* DUL and SRV are dumped.
* This parameter is used to make the transfer software display the debugging
character string on the console.
Description * This parameter does not affect DICOM data transfer.
(8) -jn
-jn
This is the time to take timeout.
Meaning
* Sets the time to take timeout in seconds.
Description * This parameter is to be changed when taking timeout.
(9) -k
-k
level = 0
1
2
Meaning
* A variety of specifications have since been needed in conjunction with DICOM
modality LUT support.
Description
* If DICOM Modality LUT OD is enabled, set appropriate options to suit each output
destination.
Argument : Data necessary for executing a function, subroutine, procedure, or other operation.
Arguments are assigned to functions and subroutines when executed. For example,
the argument in f (
χ
) is
χ
.
PDU :
Protocol data unit
The types of PDUs include get-request, get-next-request, get-response, set-request,
nd trap.
Note
For details about the -v parameter, see “Checking the Error Log”.
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......