3. CXDI-31 Imaging Unit
- 8 -
(3)
Power cable fixing method
1)
Remove the yellow tapes from the power cables.
2)
Apply the tape to the position indicated 1 to 3 below in order. Before applying the tape,
straighten the cable so that there is no slack. Do not pull on the cables at this time. When
applying the tape, press the tape surface with a
cotton swab
or so to apply it firmly. (Do
not use a tool with a thin tip such as metal
tweezers
to avoid making damage on the
cable).
Recommended tape size (unit: mm)
1
2
3
(12)
14 to 18
Criteria of Tape
Application
Common to 2, 3 and 5
Tape should cover the
inner wall of the rim.
Fit in to the bottom edge of the cover.
(Press the tape firmly to the edge using a
cotton swab or so.)
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......