2 Installation Manual
- 81 -
6) Create the graph below based on the data measured in step 5).
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
1
2
3
4
5
6
7
8
9
10
11
Step
Density
As shown in the above graph, the measurement values need only to nearly form a straight
line
from the minimum density to the maximum density.
The important point here is that the measurement values make a straight line and a
maximum
density corresponding to the settings is output.
If the measurement values deviate too far from the straight line and a maximum density
corresponding to the settings is not output, printer and monitor output linearity settings,
calibration, and other adjustments are necessary.
* The above graph is an example of measurements when the maximum density is
set to 3.2 D and the printed film is measured with a densitometer. Refer to the
data below when the maximum density is set to 3.1 D, or when gradation analysis
software is used to take measurements on the monitor.
* The LINEAR output cannot be set for some printer models.
In this case, try to select an output setting as close as possible to LINEAR.
Ideal density values in LINEAR LINE
Step
1
2
3
4
5
6
7
8
9
10
11
3.2 D
0
0.32
0.64
0.96
1.28
1.60
1.92
2.24
2.56
2.88
3.20
3.1 D
0
0.31
0.62
0.93
1.24
1.55
1.86
2.17
2.48
2.79
3.10
12 Bit
0
410
819
1229
1638
2048
2457
2867
3276
3686
4095
8 Bit
0
26
51
77
102
128
153
179
204
230
255
Possible causes of non-linearity in measurement values:
Inadequate calibration of printer and high-definition monitor
Inadequate settings for CXDI printer parameters
Faults in printer or high-definition monitor
Max setting
density
Film base
density
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......