2 Installation Manual
- 44 -
4.3.9. TABLE SETUP Settings
(1) Purpose
Adjust the CXDI operation unit's TABLE SETUP to match the exposure conditions (X-ray
tube voltage, X-ray tube current, msec or mAs value) of the X-ray generator.
(2) Procedure
1) Boot the CXDI system.
2) Open the TABLE SETUP Change window from the Normal Exposure window.
System
Æ
SETUP MENU
Æ
SYS. SETUP
Æ
TABLE SETUP
3) Select the tabs to be changed and change the X-ray tube voltage, X-ray tube current, and
msec or mAs value data to match the exposure conditions of the X-ray generator.
* See the operation manual for the details of settings.
4) After finishing the changes, return to the Normal Exposure window and check that the
TABLE SETUP has been changed.
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......