2 Installation Manual
- 69 -
Printer Model Specifications (Reference)
•
Kodak MLP190
Maxmum
equivalet area
in CXDI
2048 x 2560 (@160
µ
m)
Specifications
80
µ
m x 4096 x 5120
•
Kodak KE Kodak Print Spooler Model 100
Maxmum
equivalet area
in CXDI
2018 x 2528(@160
µ
m)
Specifications
79
µ
m x 4090 x 5120 (value after passing through the print spooler)
•
The above settings are the size of the effective area when the image passes
through the print spooler and the image is plotted up to the annotation area.
In other words, these settings do not display an annotation area, instead
handling it as an image area.
•
If the data is transferred without setting [Image creation in CXDI], the
Requested Image Size setting is used in DICOM. In this case, the
annotation area is automatically displayed in the 2180 printer. As a result, a
maximum image area of 79 um x 4090 x 4996 must be designated. In this
case, the CXDI relies on the 2180 for image rotation (Film Orientation), but
images larger than 1.7 MB cannot be rotated by the 2180. Therefore,
operation without the setting for [Image creation in CXDI] cannot be
performed in the 2180. (Although operation is possible by setting Requested
Image Size only for using DICOM without the setting for [Image creation
in CXDI], this option is not installed in the CXDI.)
•
Although the resolution of the printer itself is 79 um x 4090 x 5260, this
complete resolution cannot be used when the image passes through the
spooler.
When “_” is used in AE Title, the association is rejected.
Use the Disable function for N-EVENT-REPORT to disable this setting.
The FilmSize parameter can be used. The MediaType (BLUE, CLEAR)
parameter is also supported.
•
Although the Film Orientation parameter is supported up to 1.7 MB, in
actuality, DR images cannot be rotated. Like the DryView8700, the images
must be rotated by the CXDI side.
•
When Requested Image Size is expanded, the maximum plotting size is
limited (79 um x 4090 x 4996) so that the annotation area can be obtained.
When a Requested Image Size expansion error occurs, the image is
interpolated and printed at the suitable size. In this case, the error does not
return to the CXDI side. For example, the image is printed at
310 mm even if 326 mm is designated.
•
If an expansion error occurs in the spooler when Multi Display Format is
used, the print queue cannot be processed.
•
The system is in a critical state when a Failure status is indicated. A user
message is displayed indicating this state, and images are no longer
transferred. (Fully installed)
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......