2 Installation Manual
- 58 -
(9) -m smoothing
-m smoothing
Type of smoothing (character string)
Meaning
* This parameter is used in the DICOM Basic Film Box (2010, 0080).
* This parameter designates the smoothing method for the image.
* In the DICOM standards, this parameter setting is valid only when CUBIC is
selected for the magnification parameter above.
* In the DICOM standards, value to be transferred is not predetermined.
* The designation method varies according to the printer. For example, the
MLP190 uses -m NORMAL.
Description
Æ
The parameter is not transferred over DICOM if it is not designated. In this
case, the printer default values are used. When this parameter is not transferred,
problems can occur since unsuitable default values may be used.
* This parameter is determined by asking the printer engineer or by viewing the
conformance statement.
(10) -S configuration
-S configuration
Adjustment information (character string)
Meaning
* This parameter sets the printer (image quality) adjustment from the SCU side.
* In the DICOM standards, value to be transferred is not predetermined.
* The designation method varies according to the printer.
* This parameter is used in the DICOM Basic Film Box (2010, 0150).
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used. When this parameter is not transferred,
problems can occur since unsuitable default values may be used.
* This parameter is determined by asking the printer engineer or by viewing the
conformance statement.
(11) -O Orientation
-O Orientation
Film orientation (PORTRAIT/LANDSCAPE)
Meaning
* In versions before 2.0, printers must operate based on this parameter.
* When using image cutout from 17 x 17 inch size in the CXDI, this parameter is
set and transferred automatically.
* This parameter is used in the DICOM Basic Film Box (2010, 0040).
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used.
* Starting from version 2.0, the image can be rotated from the CXDI side without
using this parameter.
(12) -A max_density
-A max_density
Maximum density (Dx100)
Meaning
* This parameter designates the density of the digital value for 0 (4095 for
reverse display) of the CXDI transfer data image pixels. (In the CXDI, 0
indicates black.)
* In the CXDI, this parameter is used to adjust the density. Therefore, be sure to
always check that it is operating.
Description
Æ
If this parameter is not designated, it is not transferred over DICOM. In this
case, the printer default values are used. When this parameter is not transferred,
problems can occur since unsuitable default values may be used.
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......