2 Installation Manual
- 127 -
9. Specifications & Standards
9.1. Specifications
9.1.1. Imaging unit
The Imaging unit consists of a sensor, a built-in sensor power supply, an A/D board and
a cover, etc. X-rays are read electronically as image signals by the sensor, according to
changes in visible light from the fluorescent screen. Captured image signals undergo
A/D conversion and pass through the 31EO box and the multi box for transmission to a
PC.
Item Details
Remarks
Object General
shooting
X-ray
sensor
Scintillator and Amorphous
Silicon (a-Si) sensor
LANMIT sensor
Effective filming range
226mm x 288mm
Pixel matrix size
2256 x 2878
Unit pixel
No. of pixels
About 6.5 million
Pixel size
100 µm
Sensor output gradation
16,384
14 bit
Output gradation
4,096
12 bit
Interface
DICOM3.0 (Ethernet)
Via control PC
Grid type
Fixed grid
6 options
System control unit
Control PC, multi box
Sold separately
E/O box
Provided
Bundled
Sensor cable
Length: 4.5m (Dia.: 10mm)
X-ray monitor
1 location
Built in
Sensor DC-DC power supply Built in
Grit detection
Provided
Present/not present only
Position detection
None
Status display
Built in LED
Color (Green and Red)
Power consumption
Max. 45W (av. 12W)
External
dimensions
324(W) x 327(L) x
20.3(H)mm
Weight 2.8kg
(cable:
0.5kg)
[Table 1]
9.1.2. E/O box
There is a cut-off function for the film camera unit’s transmission signal and power
supply.
Transmitted signals are converted from electrical into optical signals.
The CXDI-31 can be connected to two E/O boxes.
Item Details
Remarks
External dimensions
270(W) x 190(L) x 65(H)mm
Weight 2.2
kg
[Table 2]
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......