2 Installation Manual
- 99 -
4.3.17. Tool Modes
(1) Purpose
The tool modes (startup options) are intended in order to check operation, and are used to
launch the CXDI application on the control PC by itself, and to display items that are not
normally displayed.
(2) Preparation
1) Connect the keyboard and the mouse to the control PC.
2) Delete the “ccrstart.bat” file from Startup.
3) Disconnect the imaging unit from the control PC.
(3) Notes
Perform the following operation before using the /np mode. Especially, be sure to back up
the exposure mode names and the customized settings before the operation.
1) When using /np with the same settings as that of the connected sensor unit
BodyPart**.ini file can be used as it is.
Example
Connection
/np
setting
Sensor1 Table Table
Sensor2 Stand Stand
2) When using /np with different settings from that of the connected sensor unit
Move the BodyPart**.ini file in the BodyParts folder to the desktop, etc. However, do not
move the Reference folder. If the BodyPart**.ini file is left in the BodyParts folder,
system will not be able to be started, as the sensor type of the BodyPart and the settings
do not match.
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......