2 Installation Manual
- 51 -
C. Press the “SET” button, and input the parameters.
(Normally, the parameters need not be set. They must be input only when the need
arises.)
* A space must be input between each of the parameters.
[Fig 5]
D. After setting the output destinations, follow the procedure below to check whether images
can actually be transmitted. Return to the user menu, capture a sample image (one X-ray
image), and transmit the image to the printer and storage. There are two errors that may
result if the image cannot be transmitted:
a. “DICOM Connect Error. Cannot connect to the target. Check network or port number
setting. Retry?”
b. “DICOM Transfer Error. Error occurred during the association. Retry?”
Message a indicates that connection at the TCP/IP level is not possible and that the
physical connections or the subnet mask and other settings must be checked again.
Message b indicates that communication at the TCP/IP level is problem-free but that
DICOM level communication has failed. In this case, check again that AE_TITLE of
CXDI has been sent properly to the transmission destination and that the IP address, port
number and AE_TITLE of the transmission destination which are set with CXDI have
been set properly.
* Upper-case letters are used to input “AE_TITLE” of the transmission destination.
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......