4.3.12. Linearity Check of Transfer Image Density
(1) Purpose
An SMPTE image is used to check whether the density linearity of the image printed out by
the printer and the image displayed on the high-definition monitor matches the density
linearity of the image transferred by the CXDI.
(2) Notes
1) This checking procedure should be performed before the procedure in the section of
“Operation Unit Gamma Correction”.
2) The adjustment and checking procedures below should be completed before performing this
procedure.
A) The printer and high-definition monitor connections and setting adjustments should be
completed. The printer and monitor image output settings should be set to LINEAR.
For example, if the KODAK MLP190 is connected, the printer parameter
“-S configuration” must be set to “-S CS000”. For other printers, refer to “Printers and
Parameter Examples Reference” in “Network Connections”, and set so that the curve shape
is 0 (density linear).
B) Be sure to calibrate the printer and high-definition monitor units separately before
performing this procedure.
C) When the imaging screen is “tray type”, change it to “category type” by selecting:
System
Æ
SETUP MENU
Æ
CUSTOMIZE DISPLAY.
(3)
Rough adjustment
1) Start up the CXDI system.
2) Using the two knobs at the rear of the operation unit, adjust the brightness and contrast of
the touch panel screen for optimum visibility.
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......