2 Installation Manual
- 91 -
4.3.14. Changing the Total Image Count
(1) Purpose
When the imaging unit is replaced (including the replacement of the LANMIT) for servicing,
the total image count displayed on the user screen can be returned to “0”if necessary.
(2)
Notes
1) The CXDI is connected by the system.
2) Set the CXDI application so that it does not start up.
3) Files are overwritten, and so be careful when performing the procedure.
4) As an extra precaution, write down the numerical values before overwriting them.
5) The overwritten counter becomes valid the next time that the CXDI is started up.
(3) Procedure
1) Turn on the control computer, and then start up Windows NT.
2) Open Windows NT Explorer.
Start
Æ
Programs
Æ
Windows NT
3) A file called “ExpResult.ini” is contained in the CCR folder.
Open this file. (D:\CCR\ExpResult.ini) [Fig 1]
[Fig 1]
Содержание CXDI-31
Страница 11: ...Labels and Markings Rear iii ...
Страница 14: ...Labels and Markings Rear vi ...
Страница 18: ......
Страница 20: ......
Страница 34: ...2 Installation Manual No 7 No Name E O power connector guard Name Remarks Remarks Table 3 8 ...
Страница 37: ...2 Installation Manual 4 Installation 4 1 Connecting the units 4 1 1 Connecting block diagram Fig 1 11 ...
Страница 46: ...2 Installation Manual 20 2 Set up on the left side Screw x 4 on the E O box Fig 9 Fig 10 ...
Страница 150: ...2 Installation Manual 124 8 External Dimensional Diagram 8 1 CXDI 31 8 1 1 Imaging unit Fig 1 8 1 2 E O Box Fig 2 ...
Страница 151: ...2 Installation Manual 125 8 1 3 Storage Box Fig 3 ...
Страница 152: ...2 Installation Manual 126 8 2 Option 8 2 1 Grid L 21 8mm 4 1 22 3mm 8 1 22 5mm 10 1 Fig 4 ...
Страница 157: ...3 CXDI 31 Imaging Unit 1 1 Unit Diagram Fig 1 DC DC Conversion Board ...
Страница 184: ......
Страница 186: ......
Страница 187: ...4 CXDI 31 E O Box 1 CXDI 31 E O Box Unit Diagram Fig 1 1 ...
Страница 195: ...4 CXDI 31 E O Box 2 CXDI 31 I F PCB Dip Switch SW1 Fig 5 No change SW 2 1 OFF 2 OFF 3 OFF 4 OFF Fig 6 9 ...
Страница 198: ......