D-1
Appendix D.
Digits Conversion
Vibrating-wire sensors are typically supplied with a calibration report from the
manufacturer. Use the calibration data to convert the frequency output of the
CDM-VW300 to engineering units. While basic measurement of CDM-VW300
analyzers is frequency, which is expressed in hertz (Hz), some calibration reports
express the base measurement in terms of 'digits'. If this is the case, first convert
the frequency output of the CDM-VW300 to a digits equivalent using the
following formula,
digits = Hz
2
/ 1000.
Use gauge factor and polynomial fit data from the calibration report to then
convert digits to engineering units. Following is an example that converts the
frequency output of the CDM-VW300 first to digits, then to engineering units of
millimetres of displacement.
D.1 Example: Frequency to Digits to Displacement
Measuring a Geokon vibrating-wire displacement sensor, the CDM-VW300
analyzer outputs a value of 2400 Hz. Figure D-1 shows the calibration report for
the sensor. The relationship of sensor output to displacement,
s
, is expressed as,
s = (3.598e–9) • digits
2
+ (1.202e–3) • (–3.1682)
First, 2400 Hz is converted to digits using the following formula:
digits = (2400 Hz)
2
/ 1000 = 5760
Next, the digits value is placed in the polynomial equation:
s = (9.139E–08) • (5760)
2
+ (0.03054) • 5760 + (–80.471)
s = 98.53 mm
Содержание CDM-VW300 Series
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 59: ...User Manual 47 Figure 7 16 LoggerNet connect screens showing frequencies from CDM VW300 ...
Страница 70: ...CDM VW300 Series Dynamic Vibrating Wire Analyzers 58 ...
Страница 76: ...Appendix B SC CPI Datalogger to CPI Interface B 4 ...
Страница 80: ...Appendix C CDM Devices and CPI Bus C 4 Figure C 2 Long cable lengths of a distributed CPI bus ...
Страница 82: ...Appendix D Digits Conversion D 2 Figure D 1 Geokon Calibration Report of a Sensor without a Thermistor ...
Страница 86: ...Appendix E Calculating Measurement Error E 4 ...
Страница 116: ...Appendix G CRBasic Program Library G 26 ...
Страница 117: ...Appendix G CRBasic Program Library G 27 ...