User Manual
37
Figure 7-8. Five-wire vibrating-wire sensor connection
Since temperature is a factor that can influence the frequency output of a
vibrating-wire sensor, a temperature measurement is often part of the calculation
used to convert the frequency output of the sensor to the desired engineering units.
A thermistor is a specialized resistor used for measuring temperature and as such
has no polarity.
7.7.3
CDM-VW300 to SC-CPI Connection
Connect a CPI cable between the CPI port of the CDM-VW300 and the SC-CPI
interface as shown in Figure 7-9. A standard RJ45 cable, such as those used in
Ethernet networks, may be used. Yellow tape included in the CPI Network Kit
(pn #29370) can be used to identify an RJ45 cable that is being used for CPI bus
communications.