87
Generator Type
–
The buttons labeled “Isolated” and “Gnd Ref” allow the user to select
whether the DUT being tested has an isolated output or ground referenced output. Some
generators such as Hyfrecators do not have a return pad and rely on earth ground as the
current return path. When testing these generators, select Gnd Ref and the ESU-2400
will configure the variable load to measure current from the active input to earth ground.
Configure Footswitch
–
This button opens the footswitch configuration menu that will
select the footswitch output and duration that is used when the footswitch trigger button
is pressed. To close the footswitch setup menu, press the Configure Footswitch button.