55
Power Load Curves
This mode provides for automated load curve testing of the DUT. Load curves can be
configured for multiple load values and power levels.
REM/ARM/CQM
This mode allows the user to test the REM/ARM/CQM function of the DUT. The
resistance can be manually entered, adjusted by a specific resistance, or adjusted by a
percent of resistance. This mode also has the ability to set the CQM load to an open
circuit.
Measure RF Leakage
This mode provides for RF Leakage measurements. The user can configure the ESU-
2400 for the desired RF Leakage measurement including Pulsed/Continuous waveform
measurement, load setting, etc. All switching for the leakage modes is performed by
internal relays.
System Tools
This mode provides for configuration of the ESU-2400. In this mode the user can calibrate
the touchscreen, update the system firmware, and adjust settings such as filter rate and
system volume.