61
DUT Setup
For RF Energy / Leakage and Load Curve steps, the user needs to configure how the
DUT will be activated. For these measurements, the DUT can be activated by the
Operator or the ESU-2400 as shown below. When Operator is selected, the user will be
prompted when to activate or deactivate the ESU.
Auto Advance on Pass
– This setting allows for automatically stepping through the
autosequence if a measurement passes the tolerance limits that are configured for the
selected step.
Auto Capture
–
When enabled, this setting will automatically capture and validate the
DUT output when the current step is reached during the autosequence. If unchecked,
the autosequence will show instructions to the user and the user must manually select
the meter display and capture an RF measurement.