170
LOAD BANK SPECIFICATIONS
MAXIMUM
CURRENT
0 Ω
8 A RMS
1
– 6400 Ω
3.5 A RMS
INTERNAL LOAD
SELECTION
RANGE
0-6400 Ohms
RESOLUTION
1 Ohm
ACCURACY
1% ±0.5 Ohm, Non-Inductive
POWER
RATING
1 ohm: 25W
2 ohm: 50W
4 ohm: 100W
Remaining loads: 225W
DUTY CYCLE
10 seconds on, 30 seconds off
LOAD
COOLING
Dual 120mm Variable Speed Fans
Controlled by load temperature or input
power measurement
EXTERNAL LOAD
SELECTION
RANGE
0-6400 Ohms
RESOLUTION
1 Ohm
PULSE MODE TIMING MEASUREMENT
RESOLUTION
0.1 ms
ACCURACY
± 0.2 ms