139
DFA and DFA
2
(Patent Pending) Digital Fast Acquisition Technology
®
are revolutionary
new methods of measuring ESU generator output power. A high-speed analog to digital
converter is used to digitize the high frequency, high power output of the ESU generator.
An RF Current Transformer is used to convert the current signal to a voltage signal, which
is read by the analog to digital converter. By digitizing the signal a more accurate,
frequency independent measurement can be made.
The ESU-2400H makes further advances by introducing DFA
2
(Patent Pending) Digital
Fast Acquisition Technology. The high-speed analog to digital conversion is now
processed by a programmable logic system allowing for user-selectable parameters and
the ability for the ESU-2400H to be upgraded as the state of the art progresses in
Electrosurgical Generators. This revolutionary new method also allows advanced
detection of pulsed waveforms, allowing for accurate measurement of signals with
complex pulsed waveforms.
DFA
®
TECHNOLOGY
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......