145
CONFigure
:CQM
:LOAD
OPEN or <numeric_value>
Open circuit or 0-500 (ohms)
:COHMs <numeric_value>
Range 0-500, sets the amount
to change the CQM
resistance when the ohms
UP/Down buttons are used
:UP
Same as pressing Change by
Ohms Up button
:DOWN
Same as pressing Change by
Ohms Down button
:POHMs <numeric_value>
Range 0-500, sets the amount
to change the CQM
resistance when the percent
UP/Down buttons are used
:UP
Same as pressing Change by
Percent Up button
:DOWN
Same as pressing Change by
Percent Down button
:HOLD
ON | OFF
:MODE
MAIN | RFMeasure | RFLeakage | CQM | LCURve | ASEQuence | SYStools
SYSTem Subsystem
This group allows the user to setup the startup mode for the unit, as well as directly
control the unit, as if pressing the keys on the front panel.
KEYWORD
PARAMETER FORM
COMMENTS
SYSTem
:VERsion?
Read Only
:VOLume
Range 0
– 10
:AVERaging
:FAST
<numeric_value>
Range 1-200 (samples)
:MEDium <numeric_value>
Range 1-200 (samples)
:SLOW
<numeric_value>
Range 1-200 (samples)
:WINdow <numeric_value>
Range 0.0 to 100.0 (mV)
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......