65
Load Curve Step
–
All of the options in the standard load curve are available to the load
curve step configuration.
The tabs across the top of the screen allow the user to configure the Loads, Power
settings to be used, and the Meter configuration for the Load Curve.
Setup Loads Tab
–
The loads can be configured as either a step based load change or
a list based load change. For the step based loads, the user enters the first load to be
used and the last load to be used. Then the ESU-2400 steps the loads by either a fixed
resistance or by a calculated amount to achieve a fixed number of steps for the test. In
the previous picture, the load curve would start at 50 ohms, and increment the load by 35
ohms
20
50
750
to provide 20 total steps in the test.