82
Available Parameter Descriptions
ESU-2400 Model Only
Mode
Parameter
Abbreviation
Description
P
u
lsed
a
n
d
Co
n
tin
u
o
u
s Mod
e
mV RMS
mV
This is the mV measured directly from the RF current
transformer.
mA RMS
mA
This is the converted mA measurement based on the RF
transformer mV to mA attenuation ratio.
Power in Watts
Watts
This is the computed power based on load setting and mA
measured.
mV Peak
mV Pk
This is the maximum mV measured in the buffer.
NOTE: This is shown as an absolute value.
mV Peak
– to –
Peak
mV P-P
This is the difference between maximum mV measured and
min mV measured.
mV Peak /
Peak
– to – Peak
Vp/P-P
This is the ratio of Peak divided by
Peak to Peak millivolts.
mV Positive Peak
Only
mV Pk+
This is the maximum positive mV measured in the buffer.
For asymmetric waveforms this can determine if the output
polarity is reversed.
Crest Factor
CF
This is the ratio of peak to rms of the measured waveform.
Frequency
kHz
This the frequency of the measured waveform
P
u
lsed
M
o
d
e
On
ly
Time
– Pulse On
Ton
This is the duration that the pulsed waveform is on in ms.
(See Diagram 1)
Time
– Pulse Off
Toff
This is the duration that the pulsed waveform is off in ms.
(See Diagram 1)
Time
– Total Cycle
Tcyc
This is the total cycle time of the pulsed waveform in ms.
(i.e. Ton + Toff). (See Diagram 1)
% Duty Cycle
%Duty
This is the ratio of the pulse on time (Ton)
versus cycle time (Tcyc). (See Diagram 1)
V Pulse
mV cyc
This represents the RMS mV over one pulsed cycle.
(See Diagram 1)
mA Pulse
mA cyc
This represents the RMS mA over one pulsed cycle.
(See Diagram 1)
Watts Pulse
Wcyc
This represents the RMS Watts over one pulsed cycle.
(See Diagram 1)
Ton
Toff
Tcyc
mV
mA
Watts
mVcyc
mAcyc
Wcyc
CF
Diagram 1 Pulsed mode (ESU-2400)
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...