104
The REM/ARM/CQM key from the main menu will access the Return Electrode Control
Quality Monitor Test. This menu allows the user to control the resistance between the
blue CQM terminals on the front of the ESU-2400.
CQM Resistance
– The CQM resistance can be entered directly be pressing on the
resistance indicator button. The maximum CQM resistance is 500 ohms.
Change by Ohms
– These buttons allow the user to adjust the CQM resistance up or
down by a set number of ohms. The maximum step size is 500 ohms.
Change by Percent
– These buttons allow the user to adjust the CQM resistance up or
down by a percentage of the current resistance. Maximum percent change is 500%. The
minimum resistance change is 1 ohm. That is, if the current resistance is 1 ohm and the
% change is 1%, pressing the Up / Down buttons will still change the resistance by 1 ohm.
Zero Ohms
– This button will set the CQM resistance to zero ohms.
Open Circuit
– This button will toggle the CQM resistance to an open circuit.
REM/ARM/CQM
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......