163
Question
Answer
I am testing a ForceTriad. When I
key the generator, I get a REM
alarm on the ForceTriad.
You must configure the DUT for Demo mode to bypass the REM
alarm. Use the Tool icon on the right most display and select
Demo Mode from the menu on the left most display.
The ESU-2400 shows a mV RMS
measurement. Is this the load
voltage?
No, this is the output voltage of the internal Current to Voltage
RF transformer. This was included for compatibility with the
ESU-2050 analyzer. To get load voltage, multiply the RMS
current by the load resistance.