156
Common Command Group
This subsystem provides access to common SCPI commands.
KEYWORD
COMMENTS
*IDN?
Read Only, Returns the following information:
Manufacturer, Model, Serial Number, Firmware Version, Software
Version, OS Version.
*RST
Write Only, Resets unit to default power up state. If selected, the
saved profile is loaded, see Profiles on page 115
.
*STB?
Read Only, Returns the status byte information:
Bit Value
Definition
0
1
Hold Mode
1
2
2
4
Calibration Mode
3
8
4
16
Tissue Response Mode
5
32
6
64
Tissue Response Test Complete
7
128
Tissue Response Test Started
8
256
Error Present
9
512
mV Out of Range
10
1024
mA Out of Range
11
2048
Watts Out of Range
12
4096
mV Peak Out of Range
13
8192
Crest Factor Out of Range
14 16384
Fan 1 Blocked
15 32768
Fan 2 Blocked