Load Curve Scenario 2
– I need to run a pure cut 300 Watt load curve, but there isn’t a
footswitch cable available for my DUT. I need to test my DUT at 50, 300, 500, 800, and
1500 ohms.
Solution
–
1. Select Load Curves from the main menu.
2. Select Configure Load Curve from the Load Curve Menu.
3. Press The Setup Loads tab at the top of the screen to configure the Loads
4. Select the List Based Loads option to configure the list of loads to be used for the
test.
5. Press Clear List to remove all of the previous settings from the list of loads to be
used.
6. Press the Add Load button and enter 50 in the number pad.
7. Press the Add Load button again and enter the values for the other loads to be
used; 300, 500, 800, 1500. The loads do not need to be entered in sequence, they
will automatically be sorted as they are entered. If an incorrect load is accidentally
entered, select the incorrect load by pressing on it and then press the Delete Load
button, it will be removed from the list. The screen should now look like the
following:
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......