
98
Load Curve Configuration
The tabs across the top of the screen allow the user to configure the Loads, Power, Meter
and ESU generator.
Load Config Tab
–
The loads can be configured as either a step based load change or
a list based load change. For the step based loads, the user enters the first load to be
used and the last load to be used. Then the ESU-2400 steps the loads by either a fixed
resistance or by a calculated amount to achieve a fixed number of steps for the test. In
the picture above, the load curve would start at 50 ohms, and increment the load by 35
ohms
35
50
750
to provide 20 total steps in the test.
Содержание ESU-2400
Страница 1: ...ELECTROSURGICAL UNIT ANALYZER ESU 2400 Series USER MANUAL...
Страница 2: ......
Страница 9: ...vii...
Страница 44: ...32 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 46: ...34 5 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 48: ...36 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 50: ...38 6 Trigger the DUT The ESU 2400 will show the mA RMS measurement in the display...
Страница 177: ...165 This Page Intentionally Left Blank...
Страница 186: ...174 NOTES...
Страница 187: ...175 NOTES...
Страница 188: ...176 NOTES...
Страница 189: ......